3.1.4 Min, Avg, Max, Cur values
Instrument samples current values of measured parameters from all connected
probes approximately once every second. While measuring, the instrument calculates
average and remembers the maximum and the minimum value for each parameter
(physical quantity) during the measurement. Min, Avg and Max values for all probes
are reset at the start of the measurement or when a new probe is connected to the
instrument.
One can switch between Min, Avg, Max and Cur values with the F2 ( Value ) key.
This key switches between Min, Avg, Max and Cur values for all probes connected. It
is also possible to switch only value for one probe with or key. Note that you can
switch between these values also during measurement, and not only after the
measurement is stopped.
Picture 3.6 shows a graphical presentation of the Cur, Min, Avg and Max values.
Start
Start
Start
Stop
Stop
Stop
t
t
t
Max
Actual result
Actual result
Actual result
Avg
Min
Picture 3.6: Max, Min and Avg values

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METREL Multinorm MI 6201 and is the answer not in the manual?

METREL Multinorm MI 6201 Specifications

General IconGeneral
BrandMETREL
ModelMultinorm MI 6201
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals