Multinorm/FonS/Poly
19
In the logger mode the instrument automatically stores all measurements in every
logging interval. This interval is set in Logger settings (described in chapter 5.5.2).
After the measurements are stored, the Min, Max and Avg values are reset and the
instrument starts measuring again for the next logging interval.
Picture 3.8: Logging mode – microclimatic probe connected, logging in progress
Instrument starts logging when the key is pressed. Icon from normal mode
is replaced with icon in logger mode. The number of concluded intervals is displayed
beside the LOG caption. For every logging interval the instrument stores all results that
can be measured until logging is stopped by pressing the key.
The F2 ( Value ) key switches between Min, Avg, Max and Cur values, as in
normal measuring mode. Min, Avg and Max values are cleared at the beginning of
each interval, so they are calculated for each interval separately and not for the whole
duration of logging. Elapsed time in the upper left corner of the status window also
represents elapsed time of each interval and not the whole duration of logging.
Instrument is logging until the key is pressed again or the memory is full.
Instrument can store approximately 4000 values. Table 3.1 shows how long the
instrument is logging if the memory is 100% free (cleared) with different logging
intervals.
Logging interval Estimated logging time
2 s 2 hours 15 min
10 s 11 hours
60 s 2 days 20 hours
10 min 28 days
60 min 170 days
Table 3.1: Interval and estimated logging times
If you intend to keep logging for more than just a few hours, it is recommended to
use power supply adapter to avoid discharge of the batteries.
3.2.2.2 Back to normal mode
Change the function keys to first set by pressing the key. Then press the F1
(Norm ) key to return to the normal mode. Press the key to reach Main Menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METREL Multinorm MI 6201 and is the answer not in the manual?

METREL Multinorm MI 6201 Specifications

General IconGeneral
BrandMETREL
ModelMultinorm MI 6201
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals