3.7 Result calculations
756/831 KF Coulometer, Instructions of Use
35
RS1 text RS1
RS1 decimal places 2
RS1 unit: ppm
RS1 limit control: OFF
RS1 low lim. 0.0
RS1 up lim. 0.0
RS1 L13 output: OFF
Text for result output (up to 8 characters)
Text input see page 16.
Number of decimal places for result (0...5)
Selection of result unit (ppm, mg/g, mg/ml, mg, ug,
mg/pc, %, no unit or up to 6 characters).
Limit control for the result (on, off)
The limits are checked each time a result is calculated.
If
on has been set:
Lower limit (0.0...999 999)
Upper limit (0.0...999 999)
Sets line L13 of the remote socket (OFF, active, pulse)
if the result lies outside the limits.
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00 Sample size, see page 46.
C01...C19 Method-specific operands, see page 36. They are stored with the
method in the method memory.
C21...C23 Sample specific operands, see page 46ff.
C26, 27 Mean values from silo calculations.
C30...C39 Common variables.
C40 Initial measured value of the sample.
C41 Amount of water at the end of the titration in ug.
C42 Determination time.
C43 Drift at the start of the titration.
C44 Temperature.
C45 Amount of charge in mA⋅s.