3.7 Result calculations
756/831 KF Coulometer, Instructions for Use
36
Input of method-specific operands C01...C19, key <C-FMLA>
With <C-FMLA> the operands C01...C19 can be
entered. For the calculation the operands which were
introduced in the formula are used.
The inputs are method-specific and are stored in the
method memory.
The calculation report can be printed with the key sequence
<PRINT></> (press keys repeatedly until "calc" appears in the display) <ENTER>
Operands C01 and C02
The following table gives the values for the operands C01 and C02 for the standard for-
mulas in the modes KFC and KFC-B depending on the unit in which you want the result to
be expressed and the unit in which you want to enter the sample size:
Result
in
Sample
size in
C01 C02 Result
in
Sample
size in
C01 C02
ppm
%
mg/g
g
1
1
1
1
10 000
1 000
mg/ml
ml
1
1 000
ppm
%
mg/g
mg
1 000
1
1
1
10
1
mg/ml
ul
1
1
C-FMLA
C
1

Table of Contents