4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
61
&Mode
Object Description Input range Reference
& Root
├ Mode Mode $G, $S 4.2.2.1.
: ├ .Select Mode selection KFC,KFC-B,BLANK,GLP 4.2.2.2.
├ .Name Name of current method read only/read+write 4.2.2.3.
├ .Parameter Parameters of current mode
│ ├ .CtrlPara Control parameters
│ │ ├ .EP Endpoint 0...50...±2000 4.2.2.4.
│ │ ├ .Control without meaning content, special
│ │ ├ .Content without meaning
│ │ ├ .Special Parameters for setting "special" 4.2.2.5.
│ │ │ ├ .Dyn Dynamics 0...70...2000 ditto
│ │ │ ├ .MaxRate Maximum rate 1.5...2240, max. ditto
│ │ │ ├ .MinRate Minimum rate 0.3...15...999.9, min. ditto
│ │ │ ├ .Stop Stop criterion
│ │ │ │ ├ .Type Type of stop criterion drift, rel.drift ditto
│ │ │ │ ├ .Drift Stop drift 1...5...999 ditto
│ │ │ │ ├ .RelDrift Relative stop drift 0...5...999 ditto
│ │ │ │
│ ├ .TitrPara Titration parameters
│ │ ├ .Direction Titration direction +, -, auto 4.2.2.6.
│ │ ├ .Pause Waiting time before titration 0...999 999 4.2.2.7.
│ │ ├ .ExtrT Extraction time 0...999 999 ditto
│ │ ├ .StartDrift Max.Drift for start of titration 1...20...999 4.2.2.8.
│ │ ├ .Ipol Polarization current 2, 5, 10, 20, 30 4.2.2.9.
│ │ ├ .PolElectrTest Test for polarized electrodes ON, OFF ditto
│ │ ├ .Temp Titration temperature -170.0...25.0...500.0 4.2.2.10.
│ │ ├ .TDelta Time interv. for meas.acquisition 1...2...999 999 4.2.2.11.
│ │ ├ .TMax Maximal titration time 1...999 999, OFF 4.2.2.12.
│ ├ .Statistics Statistics
│ │ ├ .Status Status of statistics calculation ON, OFF 4.2.2.13.
│ │ ├ .MeanN No. of individual determinations 2...20 ditto
│ │ ├ .ResTab Result table
│ │ │ ├ .Select original,delete n,delete all ditto
│ │ │ ├ .DelN Deletion of individual results 1...20 ditto
│ │ .Presel Preselections
│ │ ├ .Cond Conditioning ON, OFF 4.2.2.14.
│ │ ├ .DCor Drift correction
│ │ │ ├ .Type Type of drift acquisition auto, man., OFF ditto
│ │ │ ├ .Value Drift value for manual drift corr. 0.0...99.9 ditto
│ │ ├ .IReq Request of Id's after start id1, id1&2, all, OFF 4.2.2.15.
│ │ ├ .SReq Request of smpl size after start value, unit, all, OFF ditto
│ │ ├ .ReqTitr Titration during requests ON, OFF ditto
│ │ ├ .SampleUnit Unit of sample size g, 5 ASCII 4.2.2.16.
│ │ ├ .LimSmplSize Limits for sample size 4.2.2.17.
│ │ │ ├ .Status Status of limit control ON, OFF ditto
│ │ │ ├ .LoLim Lower limit 0.0...999 999 ditto
│ │ │ ├ .UpLim Upper limit 0.0...999 999 ditto
│ │ ├ .Id1Text Text instead of Id1 up to 10 ASCII char 4.2.2.18.
│ │ ├ .Id2Text Text instead of Id2 up to 10 ASCII char ditto
│ │ ├ .Id3Text Text instead of Id3 up to 10 ASCII char ditto
│ │ ├ .Cell Type of generator electrode no diaph.,diaphragm 4.2.2.19.