4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
66
&SmplData
Object Description Input range Reference
& Root
:
SmplData Sample data
: .Status Status of silo memory ON, OFF 4.2.2.54.
.OFFSilo Current sample data
.Id1 Sample identification 1 up to 12 ASCII char 4.2.2.55.
.Id2 Sample identification 2 up to 12 ASCII char ditto
.Id3 Sample identification 3 up to 12 ASCII char ditto
.ValSmpl Sample size ±X.XXXXX ditto
.UnitSmpl Unit of sample size up to 5 ASCII char ditto
.Limits Limits of sample size read only ditto
.ONSilo Current sample data
.Counter Counter of silo memory
.MaxLines Maximum lines read only 4.2.2.56.
.FirstLine First line read only ditto
.LastLine Last line read only ditto
.EditLine Editing silo lines
.1 1
st
silo line
.Method Method name up to 8 ASCII char 4.2.2.57.
.Id1 Sample identification 1 up to 12 ASCII char ditto
.Id2 Sample identification 2 up to 12 ASCII char ditto
.Id3 Sample identification 3 up to 12 ASCII char ditto
.ValSmpl Sample size ±X.XXXXX ditto
.UnitSmpl Unit of sample size up to 5 ASCII char ditto
.C24 Value of variable C24 read only ditto
.C25 Value of variable C25 read only ditto
.Mark Mark of silo line read only ditto
up to 255 lines
.DelLine Delete silo line $G 4.2.2.58.
.LineNum Line number 1...255, OFF ditto
.DelAll Delete silo line $G 4.2.2.59.
.CycleLines Cycle lines ON, OFF 4.2.2.60.
.SaveLines Save results ON, OFF 4.2.2.61.