4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
70
&Assembly
Object Description Input range Reference
& Root
:
├ Assembly Assembly control
├ . GenEl Generator electrode
│ ├ .Pulse Pulses $G 4.2.2.77.
│ │ ├ .Length Length of pulses 0...2000 ditto
│ │ ├ .Current Current of generator electrode 0, 100, 200, 400 ditto
│
├ . Meas Measuring of indicator electrode
│ ├ .Status Status ON, OFF 4.2.2.78.
│ ├ .Ipol Polarization current of electrode 2, 10, 20, 40 ditto
│
├ .Outputs I/O outputs
│ ├ .AutoEOD Automatic output of EOD ON, OFF 4.2.2.79.
│ ├ .SetLines Set I/O lines $G ditto
│ │ ├ .L0 Signal on L0 active,inactive,pulse,OFF ditto
│ │ ├ up to L13
│ ├ .ResetLines Reset I/O lines $G ditto
│
├ .Stirrer Stirrer control
│ ├ .Status Status ON, OFF 4.2.2.80.
│
├ . Bur Buret
│ ├ .Empty Empties the buret $G,$S,$H,$C 4.2.2.81.
│ ├ .Prep Prepares the buret $G,$S,$H,$C ditto
│ ├ .Rates Rates
│ │ ├ .Forward Forward rate
│ │ │ ├ .Select Type of rate control digital 4.2.2.82.
│ │ │ ├ .Digital Digital rate 0...150, max. ditto
│ │ ├ .Reverse as for forward rate
│ │ │ ├ .Select Type of rate control digital ditto
│ │ │ ├ .Digital Digital rate 0...150, max. ditto
│ ├ .Fill Fill $G,$H,$C 4.2.2.83.
│ ├ .ModeDis Dispensing $G,$S,$H,$C 4.2.2.84.
│ │ ├ .Select Type of dispensing control volume, time ditto
│ │ ├ .V Volume to be dispensed 0.0001...0.1...9999 ditto
│ │ ├ .Time Time to dispense 0.25...1...86 400 ditto
│ │ ├ .VStop Limit volume 0.0001...9999, OFF ditto
│ │ ├ .AutoFill Filling after each increment ON, OFF ditto