5.8 Remote Interface
774 Oven Sample Processor, Instructions for use
99
CTL-command
Parameter Bit Pattern Function
INIT 00000000000000 initializes the remote interface
START device1
*************1 starts device1 (for ex., Titrino, Coulometer...)
START dos1
*******1****** starts Dosimat on device1 (Titrino via "activate",
6.2141.040 cable required)
STOP device1
************1* stops device1 (for ex., Titrino, Coulometer...)
START 737
*********1**** starts Coulometer 737 (6.2141.000 cable required)
ENTER
*********1111* simulates the <ENTER> key of Titrinos or a
756 Coulometer
With the START and STOP commands a short pulse (200 ms) is emitted.
Manual stop options
Parameter Bit Pattern Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
STOP device1 ************1* stops device1 (for ex., Titrino, Coulometer...)
With the STOP command a short pulse (200 ms) is emitted.
SCN Command
Parameter Bit Pattern Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
Ready1
*******1 queries "ready" state of device1 (Titrinos. …)
End1
****1*** expects the end pulse of device1 (EOD)
Cond ok
******1* queries "Cond" state of device1 (Titrinos. …)
Cond 737
1****1** queries "Cond" state of the 737 Coulometer
no error
**1***** queries "no error" state of device1 (Titrinos. …)