2.5 Integration
774 Oven Sample Processor, Instructions for use
17
Control commands of the 774:
CTL:Rm : START Gerät1 starts Coulometer
CTL:Rm : *************1 "
CTL:RS &U.R.N loads a method in the Coulometer
CTL:RS "774BLANK"..$G here e.g. "774BLANK"
Scanning the remote lines of the 774:
SCN:Rm : **0**010 waits for 'cond. ready'
SCN:Rm : *****000 waits for end of determination
774 Oven Sample Processor — 737 Coulometer
for coulometric determinations
774
737
The 737 Coulometer is completely controlled via the remote lines.
Control commands:
CTL:Rm: *********1**** starts Coulometer with
CTL:Rm: *********0**** a start impulse
Scanning via remote lines:
SCN:Rm : 10000100 waits for 'cond. ready'
774 Oven Sample Processor — 7xx Titrino
for volumetric KF titrations
774
758
6.2141.000 cable (remote)
6.2141.020
cable (remote)