5.3 Menu organization
774 Oven Sample Processor, Instructions for use
69
parameters
>start sequence
Editor for the start sequence of a sample series
Open the submenu with <ENTER>
The sequence entered here is run once when a sample series is started.
This may be used to process blind samples for instance.
parameters
>sample sequence
Editor for processing sequence for each sample
Open the submenu with <ENTER>
This sequence is run when processing each individual sample
of a series.
parameters
>final sequence
Editor for the end sequence of a sample series
Open the submenu with <ENTER>
This sequence is run once at the end of a sample series. This may be re-
setting of the sample rack for instance.
In principle the same rules for entry are valid here as for manual operation,
i.e. after selecting a command and entering the necessary data, the entry is
accepted with <ENTER>. Afterwards the next command line is accessed
where a new command can be entered.
The LEARN mode is available for the easy entry of parameters. With certain
commands “live” values can be taken over by manual execution of a single
command. More information can be found on page 33.
Furthermore the TRACE function can be used to execute every command
line step by step. See page 34.
Navigation in a sequence is accomplished as in the other menus. In addi-
tion the <INSERT> and <DELETE> keys can be used.
<INSERT> adds a new command line above the current line in a se-
quence. It is automatically occupied by the “NOP” command that has no
function. The following lines are shifted one line downwards.
<DELETE> deletes the current line in a sequence. The lines following are
shifted one line upwards.