3
Press Alarm/ or Scale/ button to go to the display scaling
value B setting.
• The display scaling value B is indicated.
• ‘Spn and ‘D/P’ indicators turn on.
NOTE
Skip to Step 7 if the default value is acceptable.
4
Press Shift button to shift the display into the setting standby
mode.
• The fth digit starts blinking, to which you can apply changes.
5
Press Shift and Up buttons to set to ‘650.0’.
• Press Shift button to go to the next digit and Up button to change
the blinking value.
NOTE
• ‘650.0’ is a display example. Set any value within the range of -19999 to 19999.
• The decimal point position depends on the decimal point position setting. Disregard the decimal point here.
• The negative sign (-) must be set to the leftmost digit. For example, set ‘-04.00’ instead of ‘-4.00’.
6
Press Alarm/ or Scale/ button to apply the new setting.
• And the next parameter setting is indicated.
NOTE
Press Alarm/ button, and the decimal point position will be indicated.
• Press Scale/ button, and the input scaling value B will be indicated.
Blinking
Blinking