2. BASIC SETTING AND OPERATION
2.1 BASIC SETTING
This section describes ow and procedure of the basic setting.
The following shows the ow and procedure to set the input to 25 – 65 Hz and the display to 25.00 – 65.00 Hz as an exam-
ple.
2.1.1 BASIC SETTING FLOW
The basic setting is as shown in the following owchart.
2.1.2 RELATION AMONG INPUT TYPE, INPUT SCALING AND DISPLAY SCALING
The relation among input type, input scaling and display scaling is as shown in the following gure and chart.
Input type: type of input signal to 47LHZ (measuring range)
Input scaling: 0% input value (input scaling value A) and 100% input value (input scaling value B)
Display scaling: 0% display value (display scaling value A) and 100% display value (display scaling value B)
Power ON
Starting measurement
Setting decimal point position
Setting scaling
Setting input type
5.0 Hz
Input scaling value A
25.0 Hz
Display scaling value A
25.00 Hz
‘HF’
Input type
Input scaling
Display scaling
600.0 Hz
Input scaling value B
65.0 Hz
Display scaling value B
65.00 Hz
Display value
65.00 Hz
25.00 Hz
5.0 Hz 65.0 Hz
Input value
25.0 Hz