Press Alarm/ or Scale/ button to go to the advanced setting
lockout setting.
• ‘DOFF’ is indicated.
• ‘D/P’ and ‘Fnc’ indicators turn on.
3
Press Shift or Up button to select ‘D ON’.
4
Press Alarm/ or Scale/ button to apply the new setting.
• And the next parameter setting is indicated.
NOTE
Press Alarm/ button, and the loop test output lockout ‘TOFF’ or ‘T ON’ will be indicated depending on the setting.
• Press Scale/ button, and the scaling setting lockout ‘SOFF’ or ‘S ON’ will be indicated depending on the setting.
5
Hold down Alarm/ or Scale/ button for 1 second or more to return to Measuring Mode.
Blinking