Safety
4
z
Do not use the device in close proximity to the sources of strong
electromagnetic fields, such as unshielded sources of radiated emissions.
Radiated signals can reduce the accuracy of the results.
z
Ensure that the device is properly grounded, and the grounding resistance is
less than 4 Ω. Failure to do so might result in inaccurate sequencing results
or even electric shock.
z
Do not remove the device cover and expose the inner component outside.
Otherwise, it might cause electric shock.
z
Prepare the laboratory and power supply according to
Getting started on
Page 27
.
z
Ensure that the input voltage meets the device requirement.
z
Only the power cord supplied by the manufacturer can be used.
Mechanical safety
To avoid falling off and personal injury, place the device on a level surface
and ensure that the device cannot be easily moved.
Components safety
z
If the fuse blew, replace the fuse with the specified type. For details,
contact the technical support.
z
Only the software that have been provided by the manufacturer can be
installed and used on the computer. Unknown software might interfere
with normal device functions, or even cause data loss. If you need to install
antivirus software, contact the technical support in advance.
z
Do not uninstall the control software by yourself. If any problem occurs
during software operation, contact the technical support.
z
Ensure that the peripheral devices meet the IEC/EN 60950-1 standards.