Maintaining the software
59
3. Pour the waste into an appropriate waste container, and dispose
of the waste according to local regulations and safety standards of
your laboratory.
4. Add sufficient laboratory-grade water into the waste container,
attach the lid back onto the container if necessary, and gently swirl
it until all inner walls are cleaned.
5. Pour the laboratory-grade water into an appropriate waste
container.
If necessary, repeat step 4-5.
6. Clean the surface and opening of the waste container with a 75%
alcohol wipe. Ensure that no waste remains in the container.
7. Attach the lid back onto the waste container.
Maintaining the software
All software updates should be performed only by the technical
support.
Troubleshooting
If malfunction occurs during operation, the device beeps or a
message is typically displayed on the screen. Follow the prompts to
troubleshoot and solve the problem. The following table lists some
of the problems and possible solutions. If problems that are not
mentioned in this manual arise, or if you need additional assistance,
contact the technical support.
Table 2 Troubleshooting
Problem Possible cause Recommended action
After turning the
power switch to
the position, I
cannot turn on
the device.
z
The power
cord is not
plugged in.
z
Fuses blew.
1. Check whether the power cord
is plugged in.
2. Check whether the fuses blew.
3. If the problem persists, contact
the technical support.