Powering on or off the device
33
Powering on or off the device
z
Ensure that the power switch is in the position before connecting to
the power supply.
z
The mains supply socket should be a standard three-prong socket and
its protective grounding terminal should be connected to the protective
grounding cable of the power supply system. If the requirements above
are not met, the device must be protectively grounded as described in
Preparing the power supply on Page 31
.
z
Ensure that the grounding cable is connected in accordance with the
relevant standard or under the guidance of the experienced electrician.
z
We recommend that you use the power cord provided by the manufacturer
to connect to the power supply, and the power cord can be only used with
this device. Failure to do so might damage the power cord or device.
Powering on the device
Perform the following steps:
1. Connect one end of the power cord to the power port of the
device, and the other end to the mains supply.
If a UPS is prepared, connect the one end of the UPS power cord
to the device, and the other end to the mains supply.
2. Turn the power switch of the device to the position.
After you power on the device, self-test begins.
Powering off the device
Before you power off the device, ensure that the sequencing run and
wash are completed, the control software is shut down, and the flow cell
compartment door is closed. Failure to do so might damage the control
software.
Perform the following steps:
1. Tap , select Shut down. In the pop-up dialog box, select Shut
down.
2. Turn the power switch to the position.
3. Disconnect the power cord from the mains supply socket or UPS.