MGI DNBSEQ-G400 - Preparing the Laboratory; Site Requirements; Operating Environment Requirements

MGI DNBSEQ-G400
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting started
28
Preparing the laboratory
Site requirements
z
Ensure that the laboratory floor is level and with a gradient of less than
1/200, and the laboratory bench has a weight capacity of over 300 kg (661.4
lb).
z
Ensure that the laboratory is free of dust, corrosive and flammable gas, heat
and wind sources, and vibrations.
z
Ensure that the laboratory is away from direct sunlight and with good
ventilation. We recommend that you refer to the standard of a bio-safety
level 2 laboratory.
z
The device should be used with the ultra-pure water machine and
refrigerator, and can be used with the automatic library preparation system.
Therefore, enough space should be considered when placing the above
instruments.
z
Ensure that enough space is provided for ventilation, cable connection and
power switch operation.
The following figure indicates distances that are required for optimal
operation and access:
700 mm
(27.6 inches)
Device
Operation area
700 mm
(27.6 inches)
700 mm
(27.6 inches)
Figure 10 Space requirements
For space requirements of peripheral devices, refer to the relevant
manufacturer’s instructions.
Operating environment requirements
z
Temperature: 19 ℃ to 25 ℃ (66 ℉ to 77 ℉ )
z
Relative humidity: 20% RH to 80% RH, non-condensing
z
Atmospheric pressure: 70 kPa to 106 kPa