Getting started
28
Preparing the laboratory
Site requirements
z
Ensure that the laboratory floor is level and with a gradient of less than
1/200, and the laboratory bench has a weight capacity of over 300 kg (661.4
lb).
z
Ensure that the laboratory is free of dust, corrosive and flammable gas, heat
and wind sources, and vibrations.
z
Ensure that the laboratory is away from direct sunlight and with good
ventilation. We recommend that you refer to the standard of a bio-safety
level 2 laboratory.
z
The device should be used with the ultra-pure water machine and
refrigerator, and can be used with the automatic library preparation system.
Therefore, enough space should be considered when placing the above
instruments.
z
Ensure that enough space is provided for ventilation, cable connection and
power switch operation.
The following figure indicates distances that are required for optimal
operation and access:
≥ 700 mm
(27.6 inches)
Device
Operation area
≥ 700 mm
(27.6 inches)
≥ 700 mm
Figure 10 Space requirements
For space requirements of peripheral devices, refer to the relevant
manufacturer’s instructions.
Operating environment requirements
z
Temperature: 19 ℃ to 25 ℃ (66 ℉ to 77 ℉ )
z
Relative humidity: 20% RH to 80% RH, non-condensing
z
Atmospheric pressure: 70 kPa to 106 kPa