Performance specifications
69
The test results are as below:
Table 4 Test results of “genetic sequencer precision manufacturer
references”
Performance
requirement
Experiment Accuracy rate Precision rate
Accuracy:
Based on the
default operation
and analysis
process, obtain
the number of
mismatched
bases by
comparing the
sequencing
results with the
target reference
sequence.
Accuracy rate is
no less than 99%
z
Experiment 1:
99.77%
z
Experiment 2:
99.66%
z
Experiment 3:
99.74%
0.06%
Repetition:
In at least 3 times
experiments on
manufacturer
references
of genetic
sequencers, the
repeated accuracy
rate is no less
than 99%
The results show that in three experiments, the accuracy rates are
all between 99.66% and 99.77%, which comply with the required
results of the manufacturer references of genetic sequencers for
performance evaluation on the genetic sequencers (99.0%). The
reproducibility of the device is good for the same reference, and the
precision rate is 0.06%.