Removing Packing Material and Making Internal Connections SediGraph III 5120 Installation
8 Mar 2010
8. Remove X-ray shields:
a. Loosen the screw on the first section of the X-ray shield, then remove the shield.
b. Loosen the two screws on the second section of the X-ray shield, then remove the
shield.
9. Connect the high voltage line:
a. Remove the high voltage line from its holder.
b. Insert the male connector into the female connector until it snaps into place.
c. Place the high voltage line in its holder.
a b
a b c