Removing Packing Material From the Unit SediGraph III 5120 Installation
38 Mar 2010
Appendix A: MasterTech Installation
Removing Packing Material From the Unit
After removing the MasterTech from the shipping carton, remove all foam packing material
from the unit before proceeding with installation.
Selecting the Voltage
The MasterTech 052 leaves the factory set for 120 VAC and with the line fuse removed. The
correct setting of the universal power entrance must be checked and the appropriate fuse
installed before the MasterTech can be operated. The MasterTech is designed to operate with
either 100, 120, 220 or 240 VAC at 50 or 60 Hz. Voltage selection and fusing are made at the
power connector which is located on the rear panel of the unit.
1. Make sure the power cord is disconnected from the MasterTech.
2. Using a pointed object, remove the fuse block and cover assembly from the power
connector at the rear of the MasterTech.
The power cord should be disconnected from the MasterTech before removing the
cover from the power input connector. Failure to disconnect the power cord could
result in electrical shock.

Other manuals for Micromeritics SediGraph III 5120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics SediGraph III 5120 and is the answer not in the manual?

Micromeritics SediGraph III 5120 Specifications

General IconGeneral
BrandMicromeritics
ModelSediGraph III 5120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals