SediGraph III 5120 Installation Verifying Operation
Mar 2010 27
Performing a Long Scan
1. Raise the cell to about 100-1500 steps.
2. Make sure the cell if filled with water.
3. Enable the Service Test mode by selecting Options > Service Text Mode.
4. Enter the password, then click
OK.
5. Select Unit [n] > Service Test > Open.
6. Enter LSXXXXY.SVT where LS=indicates a Long Scan, XXXX= unit serial number and
Y= sequence number of the test. For example: LS01011.svt states this is a Long Scan
performed on unit number 0101 and it is the first test performed.
7. Click
OK, then click Yes to create the file and display the Service Test screen.