A Run Conditions
DEGAS CONDITIONS FILE SETTINGS
Leave these settings at the default. Do not modify.
ANALYSIS CONDITIONS FILE SETTINGS
Field Enter
Description
Run Conditions
Isotherm Collection
Target Pressure
Pressures - Dp not adjust
Options -
l Use first pressure fixed dose: No
l Use maximum volume increment: No
l Absolute pressure tolerance: 3.0000 mmHg
l Relative pressure tolerance: 3.0%
Preparations
l Fast evacuation: Yes
l Evacuation time: 0.30 hours
l Leak test: No
Free Space
l Free space group: Measured
l Lower dewar after free space: No
l Evacuation time: 0.02 hours
l Outgas test: No
P
0
and T
l P0 type: Measure
l P0: 745.000 mmHg
l Temperature: 77.30 K
Equilibration
l Equilibration interval: 6 sec
l Minimum equilibration delay at P/P0 > 0.995: 600 secs
Backfill
Do not adjust
A - 2
TriStar II 3020 Confirm Installation Instructions and Checklist
302-42870-04 (Rev A) Sep 2017