A RUN CONDITIONS
AUTOMATIC BLANK RUN CONDITIONS
The TriStar II 3020 application contains an automated procedure of conducting the blank tube run.
Use following steps to create blank tube run conditions and start the procedure.
1. Go to Unit 1 > Service Test >PCPAnalysis.
2. In the Analysis Type group box, select Nitrogen Blank.
3. Click Start.
4. When the run is complete, a report is created for each sample port. The Isotherm report con-
tains summary information about the run.
MANUAL BLANK RUN CONDITIONS
SAMPLE INFORMATION FILE SETTINGS
Field Enter
Sample
Blank Tube Test
Operator
Enter the Operator initials.
Submitter
Analyzer Serial Number
Mass
1.0000 g
Density
1.0000 cm
3
/g
Type of Data
Automatically Collected
SAMPLE TUBE FILE SETTINGS
Field Enter
SampleTube
Default 3/8 in. Sample Tube
Stem Diameter
3/8 in.
Physical volume below mark
4.8608 cm
3
User Isothermal Jacket
Yes
Use Filler Rod
Yes
A Run Conditions
TriStar II 3020 Confirm Installation Instructions and Checklist
302-42870-04 (Rev A) — Sep 2017
A - 1