INSTALL THE SAMPLE TUBE
Repeat these steps for each sample tube to be installed.
If using...
Then...
Arubber stopper
Remove it.
An isothermal jacket
Slide the jacket down over the stem of the sample tube until it touches
the sample tube bulb.
Afiller rod
Hold the sample tube horizontally and carefully slide the filler rod into the
tube.
Do not hold the rod vertically and drop the rod into the tube;
this could break the rod and/or tube.
1. Remove the sample tube stopper.
l If using a filler rod, hold the sample tube horizontally and carefully side the filler rod into the
tube.
l If using an isothermal jacket, slide the jacket down over the stem of the sample tube until it
touches the sample tube bulb.
2. Insert the P
0
tube into the center hold of the dewar cover and slide the cover up. Ensure the
tapered side of the cover is facing downward.
3. Insert the sample tube through one of the holes in the cover.
Install the Sample Tube
TriStar II 3020 Confirm Installation Instructions and Checklist