ANALYSIS CONDITIONS FILE SETTINGS
Field Enter
Description
Reference material
Isotherm Collection
Pressures
Options -
Target Pressure
Do not adjust
l Use first pressure fixed dose: No
l Use maximum volume increment: No
l Absolute pressure tolerance: 5.0000 mmHg
l Relative pressure tolerance: 5.0%
Preparations
l Fast evacuation: No
l Evacuation time: 10.0mmHg/sec
l Unrestricted evacuation from: 5.0 mmHg
l Evacuation time: 0.30 hours
l Leak test: No
Free Space
l Free space group: Measured
l Lower dewar after free space: Yes
l Evacuation time: 0.10 hours
l Outgas test: No
P
0
and T
l P0 type: Measure
l P0: 760.000 mmHg
l Temperature: 77.30 K
Carbon Reference Material Run Conditions
TriStar II 3020 Confirm Installation Instructions and Checklist
302-42870-04 (Rev A) — Sep 2017
A - 5