48
ERROR CODE DESCRIPTION
CORRECTIVE
ACTIONS
E008 Scanning error.
Please turn off the meter and
reboot it again. If the problem
recurs, please contact the
relevant technical staff.
E009
The sample may not suitable
for qLabs testing.
Please retest with a new test
strip. If the problem recurs,
please contact your local dealer
and use the laboratory method
to perform the test.
E010
The operation of sample adding
has timed out.
Make sure the sample adding
is completed before the end of
the countdown.
E011
The sample is added too early
or the strip is a used one.
Please use the new test strip
to re-test and add the sample
after the meter prompts the
user to add sample.
E012 APTT calculation error.
Please retest with a new test
strip. If the problem recurs,
please contact your local dealer
and use the laboratory method
to perform the test.
E013
The strip is defective or the
sample is added improperly.
Please retest with a new test
strip and add the sample strict-
ly as specied in the insert.
E014
The HCT of the sample is out
of range.
Please ensure that the HCT
value of the sample is between
30% and 55%.
E015 Bluetooth error.
Please turn off the meter and
reboot it. If the problem recurs,
please contact the relevant
technical staff.