ERROR CODE DESCRIPTION
CORRECTIVE
ACTIONS
E016 FIB calculation error.
Please retest with a new test
strip. If the problem recurs,
please contact your local dealer
and use the laboratory method
to perform the test.
E018 The strip is placed improperly.
Do not remove the test strip
during the test.
E019
The temperature of the meter
is too high.
Please adjust the ambient
temperature to 10 °C ~ 35
°C, and turn off the meter for
5 minutes, then turn on the
meter and restart the test. If the
problem recurs, please contact
the relevant technical staff.
E030
The meter is vibrated or
shaken.
Do not move, vibrate or shake
the meter during the test and
ensure that the meter is used in
an environment away from high
frequency vibrations.
E031 The strip is inserted inversely.
Make sure the test strip is
facing up before inserting it into
the meter.
E032 Print error.
Make sure that the meter is
properly communicating with
the base and that printing
paper at the base is installed
correctly and that there is
sufcient paper.
E033 Codechip reading error.
Please unplug the Codechip
and reinsert it, and make sure
the Codechip is inserted in
place. If the problem recurs,
please contact the relevant
technical staff.
NO COAG
No sample coagulation is
detected.
Please retest with a new test
strip. If the problem recurs,
please contact your local dealer
and use the laboratory method
to perform the test.