The following instruction Set involving frequency related settings uses 6365 as an example.
The frequency range of each model is as follows:
6363 (10Hz~20KHz) 6364 (10Hz~100KHz)
6365 (10Hz~200KHz) 6365A (0.1Hz~200KHz)
6366 (10Hz~500KHz) 6366A (0.1Hz~500KHz)
6367 (10Hz~1MHz) 6367A (0.1Hz~1MHz)
*IDN?
Identification Query.
Responses:Manufacturer, machine model, serial number, information on the firmware version.
Example:
*IDN?
Return
MICROTEST,6365,S00001,V1.607150
*TRG
(Triggers a direct measurement) Trigger one measurement without returning to the result.
Can be used for Meter Mode and Multi-Step Test Mode.
Example:
*TRG
:CAL:OPEN:ALL?
Execute open circuit calibration between 10Hz-200kHz frequency and read the execution results.
Responses:
OK (Calibration successful)
NG (Calibration failed)
Example:
:CAL:OPEN:ALL? Return OK