EasyManua.ls Logo

Microtest 6363 - System Settings

Microtest 6363
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
3.2 System Settings
Press the Menu control key to enter the Main Menu screen, and then select SYSTEM to enter
the system setting screen.
Figure 3-2-1 System Setup Screen (1/2)
1. Trigger Delay (mS): Set the trigger delay time; the setting range is 0-5000 and the unit
is mS.
2. Compare Mode Beep: Set the time for the buzzer to prompt when the compare
function is enabled in Meter Mode. The Select function key can be used to make
selections.
a. OFF: Disable prompt.
b. NG: Prompt when any one item exceeds the setting range (FAIL).
c. OK: Prompt only when all items with compare enabled are within the setting
range (PASS).
3. Multi-Step Mode Beep: Test result prompt setting when in Multi-Step Test Mode; the
Select function key can be used to make selections.
a. ON: Enable prompt.
b. OFF: Disable prompt.
4. Handler Interface: Set Handler Interface function ON/OFF; the Select function key can
be used to make selections.
a. ON: Enable the Handler Interface function.
b. OFF: Disable the Handler Interface function.
5. GPIB Address: Set the GPIB communication address; setting range: 1-30.

Table of Contents

Related product manuals