EasyManua.ls Logo

Microtest 6363 - MEAS:FREQ <Nr3|Disc; Meas:freq; MEAS:MODE <Disc

Microtest 6363
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52
:MEAS:FREQ <NR3|disc>
Set measurement frequency.
Parameters:
10-200000 (Frequency setting range 10Hz-200kHz)
MAX Set frequency as 200kHz.
MIN Set the frequency to 10Hz (6365)/0.1Hz (6365A).
Example:
:MEAS:FREQ 1000
:MEAS:FREQ 1k
:MEAS:FREQ 1kHz
:MEAS:FREQ 1E3
:MEAS:FREQ MAX
:MEAS:FREQ MIN
:MEAS:FREQ?
Query current measurement frequency.
Responses:
<NR3> scientific notation format is used for the return format.
Example:
:MEAS:FREQ? Return 1.000000E+03
:MEAS:MODE <disc>
SetMeter Mode.
Parameters:
AC Set theMeter Mode as AC.
DC Set theMeter Mode as DC.
Example:
:MEAS:MODE AC

Table of Contents