57
:MEAS:TRIG:MODE <disc>
Set trigger mode.
Parameters:
REP Repeat trigger mode.
SING Single Mode.
Example:
:MEAS:TRIG:MODE REP
:MEAS:TRIG:MODE?
Query current trigger mode.
Responses:
REPEAT Repeat trigger mode.
SINGLE Single Mode.
Example:
:MEAS:TRIG:MODE? Return REPEAT
:MEAS:VOLT <NR3|disc>
Set measurement voltage.
Parameters:
0.01~2.0 (Voltage setting range 10mV-2.0V)
MAX Set voltage as 2.0V.
MIN Set voltage as 10mV.
Example:
:MEAS:VOLT 1
:MEAS:VOLT 1V
:MEAS:VOLT 100mV
:MEAS:VOLT 1E-1
:MEAS:VOLT MAX
:MEAS:VOLT MIN