:TRIG?(for Meter Mode)
Under Meter Mode, single measurement can be triggered and the measured data and comparison
results for that test can be returned according to the display order of the current measurement
items. (Return time varies according to the settings of the measurement speed and measurement
frequency.)
※ Can be used for Meter Mode and Multi-Step Test Mode; here is the explanation when used in
Meter Mode.
Responses:
<Data 1>,<Data 2>,<Data 3>,<Data 4>,<Comp 1>,<Comp 2>,
<Comp 3>,<Comp 4>
Return Parameter Description:
<Data>
Measurement data; <NR3> scientific notation format is used for the return format.
Data 1-4 represents the measurement data of four test items. There are only values for items
with display enabled. For example, if only two test items were enabled, two data entries will
be returned; or when under DC mode, only the Rdc item can be measured, so only one set of
data will be returned.
<Comp>
Measurement data compare results; <NR1> is used for the return format. Its definitions are as
follows:
0 Compare function disabled.
1 Comparison result is OK.
2 Comparison result is NG.
Comp 1-4 represents the individual test data comparison results of the four test items. Once
the compare function is enabled for any one of the test items, the comparison results will be
returned; its order is consistent with <Data>.
EX1:
Four test items with all compare function disabled, then the retuned result is
<Data 1>,<Data 2>,<Data 3>,<Data 4>
EX2:
Four test items, in which the compare function is enabled for one or all items, then the
returned result is