88
3.9 Save the test data to a USB flash drive.
Under the multi-step test mode, it is possible to save test data in CSV format through the USB port
in the front of the machine. File paths are as follows:
The path where the 6363 saves the files is(USB:\6363_LCR\TestData)..
The path where the 6364 saves the files is(USB:\6364_LCR\TestData)..
The path where the 6365 saves the files is(USB:\6365_LCR\TestData)..
The path where the 6366 saves the files is(USB:\6366_LCR\TestData)..
The path where the 6367 saves the files is(USB:\6367_LCR\TestData)..
Before using this function, please make sure a USB flash drive is securely connected to the
machine and “Save the test data to a USB flash drive” is enabled in the system.
File format is as follows:
<File Name>,<Result>,<Step 1 Result>,<Step 1 Data>,
<Step 2 Result>,<Step 2 Data>…<Step n Result>,<Step n Data>
Return Parameter Description:
<File Name>
File Name
<Result>
Results of Multi-Step Mode; <NR1> is used for the return format. Its definitions are as follows:
0 Test interrupted and not completed.
1 Test steps were all OK.
2 There was NG during test steps.
<Step Result>
Test result judgement of each step. <NR1> is used for the return format. Its definitions are as
follows:
0 Not tested.
1 Test result is OK.
2 Test result is NG.
<Step Data>
Test values of each step. <NR3> scientific notation format is used for the return format.