:TRIG?(for Multi-Step Test Mode)
Under Multi-Step Test Mode, single measurement can be triggered and the measured data and
comparison results for that test can be returned according to the display order of the current
measurement items.
※ Can be used for Meter Mode and Multi-Step Test Mode; here is the explanation when used in
Multi-Step Test Mode.
Responses:
<Result>,<Step 1 Result>,<Step 1 Data>,<Step 2 Result>,
<Step 2 Data>…<Step n Result>,<Step n Data>
Return Parameter Description:
<Result>
Results of Multi-Step Mode; <NR1> is used for the return format. Its definitions are as follows:
0 Test interrupted and not completed.
1 Test steps were all OK.
2 There was NG during test steps.
<Step Result>
Test result judgement of each step. <NR1> is used for the return format. Its definitions are as
follows:
0 Not tested.
1 Test result is OK.
2 Test result is NG.
<Step Data>
Test values of each step. <NR3> scientific notation format is used for the return format.
Example:
:TRIG?
Return
1,1,1.000060E+02,1,3.791240E-04,1,1.000060E+02,1,9.999395E-03