Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
LOW FEED WATER PRESSURE
• Check Feedwater pressure and
rectify.
• Go to STANDBY Mode and go to
READY Mode to release any
trapped air in the Milli-Q System.
• Contact Millipore if problem
persists.
MILLI-Q RES < SP, REPLACE
QUANTUM
• The Milli-Q Water resistivity is <
set point.
• Dispense water to eliminate any
trapped air in the Milli-Q System.
• Replace the Quantum Cartridge.
MILLI-Q RES > MAX
• The Milli-Q Water resistivity is
out of measurement range.
• Contact Millipore.
MILLI-Q T < MIN
• The Milli-Q Water temperature is
out of measurement range.
• Contact Millipore.
MILLI-Q T > MAX
• The Milli-Q Water temperature is
out of measurement range.
• Contact Millipore.
MILLI-Q TOC > SP
• The TOC is > set point.
• Contact Millipore.
NO FLOW MEASURE
• The Flowmetre has measured <
0.2 Lpm during Volumetric
Dispensing mode.
• Push the Q-POD Unit Plunger all
the way down and release.
PERMEATE C < MIN
• The Permeate conductivity is out
of measurement range.
• Contact Millipore.
PERMEATE C > MAX
• The Permeate conductivity is out
of measurement range.
• Contact Millipore.
PERMEATE C > SP
• The Permeate conductivity is > set
point.
• Contact Millipore.
Continued on next page