136
Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
LOW FEED WATER PRESSURE
Check Feedwater pressure and
rectify.
Go to STANDBY Mode and go to
READY Mode to release any
trapped air in the Milli-Q System.
Contact Millipore if problem
persists.
MILLI-Q RES < SP, REPLACE
QUANTUM
The Milli-Q Water resistivity is <
set point.
Dispense water to eliminate any
trapped air in the Milli-Q System.
Replace the Quantum Cartridge.
MILLI-Q RES > MAX
The Milli-Q Water resistivity is
out of measurement range.
Contact Millipore.
MILLI-Q T < MIN
The Milli-Q Water temperature is
out of measurement range.
Contact Millipore.
MILLI-Q T > MAX
The Milli-Q Water temperature is
out of measurement range.
Contact Millipore.
MILLI-Q TOC > SP
The TOC is > set point.
Contact Millipore.
NO FLOW MEASURE
The Flowmetre has measured <
0.2 Lpm during Volumetric
Dispensing mode.
Push the Q-POD Unit Plunger all
the way down and release.
PERMEATE C < MIN
The Permeate conductivity is out
of measurement range.
Contact Millipore.
PERMEATE C > MAX
The Permeate conductivity is out
of measurement range.
Contact Millipore.
PERMEATE C > SP
The Permeate conductivity is > set
point.
Contact Millipore.
Continued on next page