Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
POD LOCKED
• The Q-POD (or E-POD) Unit
microswitch is locked.
• Push the Q-POD Unit Plunger all
the way down and release.
PROGARD PACK OUT
• The Progard Pack is not installed
correctly or it has been removed.
• The Milli-Q System stops
operating.
• Verify that the Progard Pack is
installed correctly.
• Contact Millipore if the problem
continues.
QUANTUM CARTRIDGE OUT
• The Quantum Cartridge is not
installed correctly or it has been
removed. The Milli-Q System
stops operating.
• Verify that the Quantum Cartridge
is installed correctly.
• Contact Millipore if the problem
continues.
RO FEED C < MIN
• The Feedwater conductivity is out
of measurement range.
• Contact Millipore.
RO FEED C > MAX
• The Feedwater conductivity is out
of measurement range.
• Contact Millipore.
RO FEED T < MIN
• The Feedwater temperature is out
of measurement range.
• Contact Millipore.
RO FEED T > MAX
• The Feedwater temperature is out
of measurement range.
• Contact Millipore.