Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
POD LOCKED
The Q-POD (or E-POD) Unit
microswitch is locked.
Push the Q-POD Unit Plunger all
the way down and release.
PROGARD PACK OUT
The Progard Pack is not installed
correctly or it has been removed.
The Milli-Q System stops
operating.
Verify that the Progard Pack is
installed correctly.
Contact Millipore if the problem
continues.
QUANTUM CARTRIDGE OUT
The Quantum Cartridge is not
installed correctly or it has been
removed. The Milli-Q System
stops operating.
Verify that the Quantum Cartridge
is installed correctly.
Contact Millipore if the problem
continues.
RO FEED C < MIN
The Feedwater conductivity is out
of measurement range.
Contact Millipore.
RO FEED C > MAX
The Feedwater conductivity is out
of measurement range.
Contact Millipore.
RO FEED T < MIN
The Feedwater temperature is out
of measurement range.
Contact Millipore.
RO FEED T > MAX
The Feedwater temperature is out
of measurement range.
Contact Millipore.