Image Optimization 5-25
Invert
Description
This function is used to set the display manner of spectrum.
Operations
Turn on or off the function through the [Invert] item on the touch screen.
Select ―Auto Invert‖ in the ―[Setup] (F10) → [System] → [Image]‖, thus the
spectrum can automatically invert when the color flow is steered to a certain angle
to accommodate operators habit of distinguishing flow direction.
Speed
Description
This function is used to set the scanning speed of PW mode imaging.
Operations
Rotate the knob under [Speed] item on the touch screen.
There are 6 levels of scan speed available, the smaller the value the faster the
speed.
Effects
Changing the speed makes it easier to identify the cardiac cycles and to detect
more details.
T/F Res
Description
This function is used to adjust for a balance between time resolution and spatial
resolution.
Operations
Adjust through the [T/F Res] item on the touch screen.
There are 4 levels of T/F Res values available.
Wall Filter
Description
It filters out low-velocity signals to provide effective information, and this function is
used to adjust the filtered frequency. The real-time value is displayed in the image
parameter area in the upper right corner of the screen.
Operations
Select through the [WF] item on the touch screen.
There are 9 levels of wall filter function provided.
Impacts
Signals of low-velocity flow may be missing.
Tint Map
Description
This function provides an imaging process based on color difference rather than
gray distinction.
Operations
Rotate the knob under the [Tint Map] item on the touch screen to select the map.
There are 8 color effect maps available.
Press the knob under the [Tint Map] item on the touch screen to turn on/off the
function.
Trace
Trace Smooth
To set the smooth level of tracing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%