5-94 Image Optimization
Map
Description
Used to adjust the color map to achieve the switch between the gray map and the
color map.
Operation
Rotate the knob under the [Map] on the touch screen to select the map.E1 is a gray
map; E2 and E3 are color maps.
Display Format
Description
Used to adjust the display format of ultrasound image and the Elasto image, and
return to the previous state.
Operation
Tap each soft key on the touch screen to complete the adjustment.
The system provides 3 types of display format:
H 1:1; V 1:1; Full.
Impacts
More accurate result is obtained based on the actual situation.
HQElasto
Description
Turn on high-quality scanning mode to optimize penetration.
Operation
Tap [HQElasto] to enable the function and the button will be highlighted.
Impacts
When the feature is activated, the system turns into single-frame scanning mode.
Pressing <Update> acquires one frame image of B and Elasto.
RLB Map
Description
Used to help the user judge the region where it is suitable for elasto measurements
based on the distinctive colors in elasto image in quality map.
The RLB Index shows the signal reliability of the STE inside the ROI, and it helps
the user judge the effectiveness of the current elasto measurement.
Operation
Tap [RLB Map] to enable the function (the soft key is highlighted). The quality map
appears on the left side of the image.
With both real-time ultrasound image and elasto image displaying, the real-time
ultrasound image is replaced with the quality image.
With only elasto image displaying, the elasto image is replaced with the quality
image.
The higher the RLB Index is, the more reliability the signal becomes; vice versa.
There are three colors showing the RLB Index: Red (low reliability), Yellow (medium
reliability) and White (high reliability).
Impacts
The color in High area is suitable for the imaging of shear wave elastography.
The color in Low area is not suitable for the imaging of shear wave elastography (liquid,
gas, bone, etc). It can be used as a tool to detect the tissue or the lesion.
RLB View
Description
This feature supports the less qualified region or the region that fails to receive
shear waves of the gas or liquid to be displayed in hollow.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%