DICOM/HL7 11-5
Tips:
If the server software supports the compression algorithm, select JPEG, RLE, JPEG2000.
Otherwise, original data should be used (RLE is the default method).
RLE, JPEG and JPEG2000 are not supported by all SCPs. Please refer to the electronic file
DICOM CONFORMANCE STATEMENT of the SCP to check whether SCP supports or not.
Do not select these compression modes if the storage server doesn’t support them.
11.1.3.2 Print Service Preset
1. On DICOM Service screen, click [Print] page tab to enter the Print page.
2. Select device, enter the right AE Title, port, etc.
3. Click [Add] to add the service to the Service List.
DICOM print preset items are described as follows:
Name
Description
Configure
New
Service
Device
After you set the server (s) in DICOM Preset screen, the name
(s) will appear in the drop-down list, select the name of the print
server.
Service Name
Default is xxx-Print, user-changeable.
AE Title
Application Entity title, here, it should be consistent with that of
the print server.
Port
DICOM communication port, 104 is default. Here, the port
should be consistent with that of the print server port.
Maximum Retries
Set the maximum retries.
Interval Time (s)
Interval time that the system retries a connection.
Timeout
Refers to timeout during association establishment.
Print
Properties
Copies
Refer to copies of printed files. You can select among 1 through
5, or directly enter the numeral.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%