EasyManuals Logo

Mindray DC-80 User Manual

Mindray DC-80
383 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
5-8 Image Optimization
iClear
Description
The function is used to enhance the image profile so as to distinguish the image
boundary for optimization.
Operations
Adjust through the [iClear] item on the touch screen.
The system provides 7 levels of iClear adjustment, off represents no iClear effect,
and the bigger the value the stronger the effect.
Impacts
It may cause noise increasing.
Persistence
Description
This function is used to superimpose and average adjacent B images, so as to
optimize the image and remove noise.
Operations
Rotate the knob under the [Persistence] item on the touch screen to adjust the
value.
The system provides 7 levels of frame average adjustment, the bigger the value
the stronger the effect.
Effects
Persistence can remove image noise to make details to be clearer.
Impacts
Persistence increasing may lead to signal missing.
Rotation /Invert (U/D Flip and L/R Flip)
Description
The function provides a better observation for image display.
Rotation
Rotate the image through the [Rotation] item on the touch screen.
Image can be rotated in the manners of angles of 0°, 90°, 180° and 270°.
When the image is rotated in the angle of 90° or 270°, the depth scale is displayed
on the upper part of the screen.
Invert (U/D
Flip and L/R
Flip)
To invert the image horizontally or vertically.
Touch [U/D Flip] or [L/R Flip] on the touch screen to invert the image
Effects
You can identify image orientation through the ―M‖ mark on the screen; by default,
the M mark is located in the upper left corner of the imaging area.
iBeam (Spatial Compound Imaging)
Description
This function is used to superimpose and average images of different steer angles
to obtain image optimization.
Operations
Adjust through the [iBeam] item on the touch screen.
The system provides 4 levels of iBeam adjustment in single B mode; Off
represents no iBeam.
Effects
Images can be optimized with less spot noise and higher resolution, so that more
details for the structure are revealed.
Impacts
iBeam is valid for linear and convex probes, and it is not available when ExFOV
function is turned on.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%