EasyManua.ls Logo

Mindray Passport V - Verification Procedures

Mindray Passport V
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verification Calibration Procedure
5 - 12 0070-10-0705 Passport V™ Service Manual
5.5 Verification
5.5.1 Initial Set-up
1. Using a patient simulator, connect the ECG, IBP1, IBP2 and temperature cables to the
left side connector panel. Set the ECG simulator for 60 bpms, 1mv QRS signal.
2. Set up the Patient Menu for Adult (Patient Size) as follows:
A. Monitor Setup
1. Display Setup – 3 Waveforms
2. ECG Speed – 25 mm/sec
3. IBP Speed – 25 mm/sec (optional)
4. Respiration / Gas Speed – 12.5 mm/sec
B. Print Setup
1. Waveform 1 ECG 1
2. Waveform 2 ECG 2
3. Select Printer – Local
C. Parameters
1. ECG
a. ECG 1 – II
b. ECG 2 – I
c. ECG 3 – III
d. ECG 1 thru 6 Size - 1 cm/mV
e. ECG 1 - I
f. ECG 2 - II
g. ECG 3 - III
h. ECG 4 - AVR
i. ECG 5 - AVL
j. ECG 6 – AVF
k. ECG 7 – V
2. NIBP
a. Set Start Pressure – 180 mmHg
Interval – 5 minutes
b. IBP1 – Scale 0 to 160 mmHg
c. IBP2 – Scale 0 to 80 mmHg
3. SpO
2
a. Averaging mode 2
b. Sensor Off Audio – off
4. CO
2
(optional)
a. Apnea Delay – 60
b. Scale 40 mmHg

Table of Contents

Other manuals for Mindray Passport V