Services Calibration Procedure
5 - 8 0070-10-0705 Passport V™ Service Manual
FIGURE 5-7 Microstream CO
2
 Calibrate Menu
5.4.5 Gas Calibration
It is recommended that to maintain accuracy of the Gas Module 3, calibration is required 
once a year or whenever gas readings appear suspicious.
NOTE: Single gas calibration is intended for 5%CO
2
 or 55%O
2
 or 
33%N
2
O or 2%Des only. For mixed gas calibration, 5% CO
2
, 
55%O
2
, 33% N
2
O and 2% Des are calibrated 
simultaneously.
NOTE: The Gas Module 3 must fully warm up before performing a 
gas calibration. For maximum accuracy, a warm-up time of 
10 minutes is recommended.
1. Use the Navigator
TM
 Knob to select Calibrate Gas.
2. Connect the gas bottle and the tubing using a T-connector as shown in FIGURE 5-8.
3. Check the airway and make sure that there is no occlusion or leakage.
4. Press the Gas Selection button to select the type of calibration gas.
5. Select Start to activate a calibration. At the start of the calibration, the message 
“Zeroing…” will be displayed for each of the gas labels as the Gas Module zeros the 
gas channels. After zeroing is successful, the Gas Module will request the calibration 
gas as indicated in the next step.
NOTE: After the calibration is started, the module will begin 
zeroing, sampling, and calibrating automatically.
NOTE: If the Gas Module fails zeroing, a zeroing error will be 
displayed and the previous calibration data will be 
restored. If so, repeat the calibration procedure from step 1. 
If problems persist, contact Customer Support.