5-4 T1 Service Manual
5.3 Performance Tests
Performance test are designed to ensure that measurement results are accurate. The following sections provide a list of
performance and accuracy tests and their recommended frequencies.
5.3.1 Performance Test Frequencies
Visual Inspection 1. When first installed or reinstalled.
ECG test Performance test 1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of relevant module.
3. Once a year for NIBP and CO
2
tests.
4. Once every two years for other parameter module performance tests.
Verification
Resp performance test
SpO
2
test
NIBP test Pressure check
Leakage test
Temp test
IBP performance test
CO
2
tests and
calibration
Leakage test
Performance test
Calibration
Analog output performance test If the user suspects that the analog output does not work.
5.3.2 Visual Inspection
Inspect the equipment for obvious signs of damage. Follow these guidelines when inspecting the equipment:
Carefully inspect the case, display screen, buttons and knob for obvious signs of damage.
Inspect the power cord, and module accessories for obvious signs of damage
Inspect all external connections for loose connectors, bent pins or frayed cables.
Make sure that safety labels and data plates on the equipment are clearly legible.

Table of Contents

Other manuals for Mindray T1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray T1 and is the answer not in the manual?

Mindray T1 Specifications

General IconGeneral
BrandMindray
ModelT1
CategoryMedical Equipment
LanguageEnglish

Related product manuals