4-2 T1 Service Manual
4.1.2 Hardware Upgrade Method
4.1.2.1 Upgrading the Monitor to Have IBP Function
When upgrading the monitor to have IBP monitoring function, please use the parameter board and parameter connector
panel assembly in the upgrade package to replace the original parameter board and parameter connector panel assembly.
To upgrade the monitor:
1. Refer to 7.3.2.2 Disassembling the parameter board and 7.3.3.2 Removing the Parameter Connector Panel Assembly
to remove the parameter board and parameter connector panel assembly.
2. Use the parts in the upgrade package to replace the original parameter board and parameter connector panel
assembly.
3. Reassemble the monitor.
4. Turn on the monitor, and then select [Main Menu][Maintenance>>][Factory Maintenance>>]enter the
required password[Device Configuration], and then select[IBP1/IBP2].
5 Refer to 5.3.8 IBP Performance Test to test the monitor.
4.1.2.2 Upgrade the Monitor to Have 12-lead ECG Function
Refer to 4.1.2.1 Upgrading the Monitor to Have IBP Function, use the parameter board in the upgrade package to replace
the original parameter board.
4.1.2.3 Upgrade the Monitor to Have 12-lead ECG Function and IBP monitoring Function
Refer to 4.1.2.1 Upgrading the Monitor to Have IBP Function, use the parameter board and parameter connector panel
assembly in the upgrade package to replace the original parameter board and parameter connector panel assembly.
4.1.2.4 Upgrade the Monitor to Have Wi-Fi Function
When upgrading the monitor to have Wi-Fi function, you need to install the Wi-Fi module (Cyberlink module PCBA or 5GHz
module) and the antenna into the upgrade package into the monitor. To upgrade the monitor:
1. Refer to 7.3.3 Disassembling the Front Housing Assembly to separate the front housing assembly and the main frame
assembly.
2. Insert the antenna in the antenna slot in the main frame, and then apply the protective tape onto the antenna.
The antenna should be inserted into the antenna slot as indicated in the following picture. Protect the antenna from
being damaged during assembly. The antenna should not bend or twist in the slot. When the antenna is in place, press
the protective tape down firmly to make sure the antenna is secured onto the main frame.

Table of Contents

Other manuals for Mindray T1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray T1 and is the answer not in the manual?

Mindray T1 Specifications

General IconGeneral
BrandMindray
ModelT1
CategoryMedical Equipment
LanguageEnglish

Related product manuals