<Setting the Manual Select Calibration>
This sets the manual select calibration function which allows the user to select a calibration channel
whose calibration data is to be used to correct indicated values.
When setting or canceling manual select calibration, you must set the color difference target
color again.
The color difference target color is not corrected using user calibration performed after mea-
surement.
See P.145 for details on manual select calibration.
Manual select calibraion cannot be set if white calibration and selected user are not done ahead of
time.
[Operating Procedure]
1
Press the Option key
Option
4
GHI
while in the measurement
screen.
The option selection screen appears.
2
Select MALTI CAL. using the key.
Selection screen
[OPTION]
 TOLERANCE
 AUTO MEAS.
 DATE & TIME
 IMPORT
 MALTI CAL.
P00  
C

P

 [0001]
[OPTION]
 TOLERANCE
 AUTO MEAS.
 DATE & TIME
 IMPORT
 MALTI CAL.
P00  
C

P

 [0001]
2,4
3,5
1