Data Processor
Functions
Selecting the limit value in P.115 displays the [limit value] selection screen.
[Operating Procedure]
1
Use the key to select a color difference target
color.
2
Press the Measure/Enter key
.
The tolerance type screen appears.
3
Select the BOX & E* as the tolerance type.
4
Press the Measure/Enter key
.
The limit value setting screen will appear.
If color difference tolerance and E are already set, go to step
6 as the set color difference tolerance data is displayed.
5,7
1,3,5,7
2,4,6,8
[TOLERANCE]  1/1
◎TARGET
 T03
 T02
 T01
P00  
C

P

 [0001]
[TOLERANCE]
◎TOLERANCE TYPE
 
 ELLIPTICAL
 BOX
 ΔE*
 BOX & ΔE*
P00   C P  [0001]
[TOLERANCE]
◎TOLERANCE
A      B
 + 0.00 + 0.00
 − 0.00 − 0.00
C      D
 + 0.00   0.00 
 − 0.00
P00   C   [0001]
P
ΔL*
Δa*
Δb*
ΔE*
Box tolerance and E
This judges both box tolerance and E.
Selection screen
Tolerance type screen
Setting screen