RDS-32 FUNCTIONALITY
18
To set the desired alarm level it may be necessary to switch between unit multipliers. To switch
between µSv/h, mSv/h or Sv/h, navigate to the unit (furthest right on the display) and change
the selected unit multiplier with the up/down arrow keys. When the desired alarm level is
selected, save by pressing the Select key. Note that Sv/h is only available in RDS-32WR
versions.
To ignore the changes, press the left arrow key until the display returns to the previous menu
level. Also waiting 12 seconds time out will switch the display back to Main display without
changing the alarm level.
In dual mode the user can set two levels which will trigger the device alarming functions: a
warning level and an alarm level.
To set the device in dual mode navigate into DR Mode and select Dual from the menu. To set
a dose rate warning and alarm navigate into Dose Rate. The first available setting will be Dose
Rate Warning. Set a suitable level, lower than the intended alarm level, according to the
process described above. Save the desired level by pressing the Select key. The setting will
move automatically to Dose Rate Alarm. Set a suitable dose rate alarm according to the
process described above.