5. FIRST START-UP
17
FILLING DISTILLED
WATER
Manual filling
Automatic filling
The first time the sterilizer is used, and later when the MIN water level indicator comes on, you
will have to fill, or top-off, the internal distilled water tank.
With reference to the figure (and with the door open), proceed as follows:
1. Fill the manual container (2 l) with distilled water,
keeping it horizontal;
2. Connect the tube’s quick connector to the
corresponding female connector under the chamber
entrance (marked
), pushing until you hear a
click;
3. Place the container in a vertical position, at the same
time, loosening the plug and taking care not
to spill
water on the machine.
4. The water will begin to flow into the tank;
5. Continue filling until the MIN level indicator turns off.
6. Continue until the water is drained from the container;
7. At this point, lower the connector below
the connection point, keeping it horizontal;
8. While pinching the tube with your fingers, press the metal lever located on the side of the
connector and detach the quick connector;
9. Refill the container (2 l) and repeat the operations described in points 2, 3 and 4 a second
time;
10. When the MAX level
icon comes on (accompanied by a beep), stop filling and detach the
quick connector as described in points 7 and 8.
NOTE
THE ICON MAX DOES NOT HAVE TO BE ON TO START A STERILIZATION PROGRAM. THE
ICON
MIN INDICATOR OFF IS SUFFICIENT.
In the event of sterilizer installation for automatic filling from an external tank or demineralizer
Milldrop (see the Chapter, “Installation”), the filling will occur automatically after the automatic
filling option has been selected.
Obviously, for the correct operation, the user must fill the external tank or switch on the
Milldrop in advance.
NOTE
USE ONLY HIGH QUALITY DISTILLED WATER. FOR THE SPECIFICATIONS OF THE WATER
SUPPLY
, SEE APPENDIX A (TECHNICAL CHARACTERISTICS).
To set the automatic filling option, please refer to the Chapter, “Configuring the Device”.
WARNING
THE AUTOMATICALLY FILLING SYSTEM MUST NEVER RUN DRY; THIS
CAUSES PREMATURE WEAR TO THE AUXILIARY WATER-INJECTION
PUMP. PERIODICALLY CHECK THE WATER LEVEL IN THE EXTERNAL
TANK.