APPENDIX A – TECHNICAL CHARACTERISTICS
55
WATER SUPPLY CHARACTERISTICS
DESCRIPTION WATER SUPPLY VALUES
VALUES IN
CONDENSATE
DRY RESIDUE
< 10 mg/l
< 1 mg/l
SILICON OXIDE SiO
2
< 1 mg/l
< 0.1 mg/l
IRON
< 0.2 mg/l
< 0.1 mg/l
CADMIUM
< 0.005 mg/l
< 0.005 mg/l
LEAD
< 0.05 mg/l
< 0.05 mg/l
HEAVY METAL RESIDUES
(except iron, cadmium and lead)
< 0.1 mg/l
< 0.1 mg/l
CHLORINES
< 2 mg/l
< 0.1 mg/l
PHOSPHATES
< 0.5 mg/l
< 0.1 mg/l
CONDUCTIVITY AT 20 °C
< 15 µs/cm < 3
µ
s/cm
pH VALUE
5 - 7
5 - 7
APPEARANCE
colorless, transparent,
without sediments
colorless, transparent,
without sediments
HARDNESS
< 0.02 mmol/l
< 0.02 mmol/l
NOTE
WHEN PURCHASING DISTILLED WATER, ALWAYS CHECK THAT THE QUALITY AND CHARACTERISTICS
DECLARED BY THE PRODUCER ARE COMPATIBLE WITH THOSE SHOWN IN THE TABLE
.
WARNING
THE USE OF WATER FOR GENERATING STEAM CONTAINING CONTAMINANTS IN
LEVELS EXCEEDING THOSE SHOWN IN THE TABLE WILL SIGNIFICANTLY SHORTEN
THE STERILIZER'S LIFE.
IN ADDITION, THIS MAY INCREASE THE OXIDATION OF MORE SENSITIVE MATERIALS
AND INCREASE LIME RESIDUES ON THE GENERATOR, BOILER, INTERNAL SUPPORTS
AND INSTRUMENTS.