6. CONFIGURATION
33
When INTERNAL DRAIN is enabled, the reading of the MAX level sensor in the internal tank
is enabled.
The EXTERNAL DRAIN command also activates the MAX level sensor located in the external
tank.
NOTE
THE LEVEL SENSOR IN THE INTERNAL TANK REMAINS ACTIVE IN ANY CASE, TO PREVENT
A POSSIBLE MALFUNCTION OF THE EXTERNAL TANK OR A MISSING OR FAULTY
CONNECTION OF THE OPTIONAL EXTERNAL DRAIN TANK
.
I
N THE CASE OF AN INSTALLATION WITH THE DRAIN CONNECTED TO THE CENTRAL
SYSTEM
, SELECT INTERNAL DRAIN.
Scroll through the items with the + and - keys; confirm with the key.
Acquisition of the
ambient pressure
(AMBIENT PRESSURE
on the SPECIAL menu)
The first time the sterilizer is used and after any reinstallation, the sterilizer must acquire the
ambient pressure.
This operation is necessary
or the correct operation of several of the device's auxiliary
systems.
When AMBIENT PRESSURE is activated, the following screen appears:
ACQUISITION OF THE
AMBIENT PRESSURE
to enter
Ç to exit
NOTE
CHECK THAT THE STERILIZER DOOR IS COMPLETELY OPEN. IF YOU TRY TO ACQUIRE
THE PRESSURE WITH THE DOOR CLOSED
THE FOLLOWING MESSAGE WILL BE DISPLAYED:
OPEN THE DOOR
TO CONTINUE
which remains until the door is opened.
Confirm the acquisition of the data by pressing the key. This message appears:
PRESSURE VALUE
SET INTO MEMORY
accompanied by a beep. The ambient data pressure has been acquired.
On the other hand, press the
Ç key to cancel the operation.